4.2 Control Ranges and Increments

NOTE:
Refer to the applicable BiPAP Vision Clinical Manual for PAV/T information.
4.2.1 Parameters
*With optional Oxygen Module
4.2.2 Alarms (Adjustable)
Alarm
High Pressure
Low Pressure
Low Pressure Delay
Apnea
Low Minute
Ventilation*
High Rate*
Low Rate*
*With optional Alarm Module
Chapter 4: Specifications and Control Ranges
Control
Range
5 to 50 cm H
0
2
Disabled to 40 cm H
0
2
0 to 60 sec.
Disabled; 20 to 40 sec.
Disabled to 99 L/min.
4 to 120 BPM
4 to 120 BPM
BiPAP Vision Service Manual
Control
Increments
1 cm H
0
2
1 cm H
0
2
1 cm H
0
2
4 set points;
Disabled, 20, 30, 40 sec.
1 L/min.
1 BPM
1 BPM