Cautions - Respironics BiPAP Vision Service Manual

Ventilatory support system
Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

2-3
Chapter 2: Warnings, Cautions, and Notes
Warnings (Continued)
2.1.3 Service
CAUTION: Electronic components used in this device are subject to damage from static
electricity. Repairs made to this device must be performed only in an antistatic,
ESD-protected environment.
Do not attempt to make connection to the diagnostic RS232 connector on the back panel of the
BiPAP Vision to obtain repair information while the unit is operating on a patient.
To assure the safety of the service technician and specified performance of the device,
Respironics recommends that only qualified technicians perform repairs to the BiPAP
Vision. Contact Respironics Technical Service for service training and authorization
information.
High voltages are present inside this device. To avoid electrical shock, disconnect the
electrical supply before attempting any repairs on the device.
For continued protection against risk of fire, replace fuses with those of the same type
and rating only.
2.1.4 Cleaning
To avoid electrical shock, unplug the BiPAP Vision unit before cleaning it.

2.2 CAUTIONS

While cleaning the unit, do not allow any liquid to enter the cabinet or the inlet filter.
Care should be taken to avoid exposing the BiPAP Vision to operating, storage, and
transport temperatures near the extremes specified in Chapter 4. If exposed to such
temperatures, allow the unit to cool or warm to room temperature before turning it on.
The unit must be positioned on its base for proper operation.
Always use an inlet filter when the BiPAP Vision is in use.
If using the oxygen module, do not exceed 100 psig oxygen supply pressure.
1015010
BiPAP Vision Service Manual