Respironics BiPAP Vision Service Manual page 180

Ventilatory support system
Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

8-38
Chapter 8: Testing and Calibration
Step 3
In Connect To, select COM1 in Connect Using, then click OK.
Step 4
In COM1 Properties, Port Settings, set the following:
Step 5
Click OK to activate the settings. At this point, a blinking cursor will appear on the screen.
Step 6
If necessary, check the ASCII settings. Go to File, Properties, Settings, ASCII Setup.
Step 7
Enter the appropriate command for the desired selection below. No typed letters will appear on the PC
or Laptop. Data will appear after the code is entered.
Step 8
If screen rolling go to: File, Properties, Settings, Emulation, VT52, OK
Note: For an existing Vent Inop condition, the MC and DC will output full screens of information. The PC
only outputs one line of error code information. No command and needs to be entered. During
normal operation, the full command needs to be entered.
B. Microsoft Windows
Step 1
From Program Manager, open Accessories.
Step 2
From Accessories, open Terminal.
Step 3
Under Settings, open Terminal Emulation then select DEC VT-52, and click OK.
Step 4
Again under Settings, open Terminal Preferences then set the following:
Step 5
Click OK
1015010
Bits per Second; 19,200
Data Bits; 8
Parity; None
Stop Bits; 1
Flow Control; Xon / Xoff.
®
V. 3.1
Line Wrap; On
Columns; 80
Cursor; Block, Blink
Show Scroll Bars
Use function arrows and control (Ctrl) keys for Windows.
BiPAP Vision Service Manual