Respironics BiPAP Vision Service Manual page 176

Ventilatory support system
Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

8-34
Chapter 8: Testing and Calibration
Option 2 Test (Nurse Call, normally closed setting)
1.
JP1 on the MC board is set at 2,3. JP2 is set at 1,2.
2.
Using a DVM and the unit turned off, measure the continuity across the High and Low of the Nurse Call/
Remote Alarm output connector. A closed circuit should be measured.
3.
Place the test orifice on the unit. Turn the Vision on and activate any patient alarm (For example, the Apnea
alarm could be set to 20 seconds).
4.
Measure the continuity on the same connector and it now should be an open circuit.
5.
Turn unit off.
Option 3 Test (Nurse Call, normally open setting)
1.
JP1 and JP2 on the MC board are set to 1,2.
2.
Using a DVM and the unit turned off, measure the continuity across the High and Low of the Nurse Call/
Remote Alarm output connector. An open circuit should be measured.
3.
Place the test orifice on the unit. Turn the Vision on and activate any patient alarm (For example, the Apnea
alarm could be set to 20 seconds).
4.
Measure the continuity on the same connector and it now should be a closed circuit.
5.
Turn unit off.
Upon completion, set the unit to the desired option.
Q.
Final Test Data Sheet
Step 1
Sign and date completed Final Test Data Sheet.
1015010
BiPAP Vision Service Manual
System Final Test (Continued)