NOTE:
Place one hand underneath the front panel enclosure while applying any pressure, so that the curve in
the front panel is not damaged. Once on and aligned properly, apply a rotating, rubbing pressure to
secure the touch pad.
NOTE:
If an alignment problem occurs, use a small amount of isopropyl alcohol to remove the touch pad. Allow
to dry before trying again.
Step 5
Installing the D / CS - LCD Mounting Plate
a.
While placing the D / CS - LCD mounting plate on the front enclosure, ensure that the
cables are properly routed. Secure the plate into position using the six mounting screws
Step 6
Installing the D / CS - Touch Pad Ribbon Cable
a.
With the tabs fully extended on (J6), place the ribbon cable completely into the connector. While
applying slight inward pressure on the ribbon cable, lock the tabs on both sides of (J6).
BiPAP Vision Service Manual
Chapter 7: Repair and Replacement
7-63