Chapter 8: Testing and Calibration
System Final Test (Continued)
L.
PAV / T mode (if installed)
Step 1
Select Mode.
Step 2
Select PAV/T.
Step 3
Activate the PAV/T mode and verify the mode features are displayed on the screen and that
no errors or alarms occur.
NOTE:
No further testing of the PAV/T mode is needed since the functioning of the PAV/T
mode is directly linked to the S/T mode through software. The S/T specifications
were verified earlier in the system final test.
Step 4
Turn off the unit, disconnect the test circuit, and remove the power.
Step 5
Install the top enclosure.
M.
Earth Resistance
Step 1
Measure and record the earth resistance value. Test current is 25 amps. The value must be
less than 0.10 Ohms (W) to pass.
N.
Earth Leakage Current
Step 1
Measure and record the Normal Pole, No Earth, L2 earth leakage current. The value must
be less than 300 microamps to pass.
Step 2
Measure and record the Reverse Pole, No Earth, L2 earth leakage current. The value must
be less than 300 microamps to pass.
Step 3
Measure and record the Reverse Pole, No Earth, No L2 earth leakage current. The value must be less
than 1,000 microamps to pass.
Step 4
Measure and record the Normal Pole, No Earth, No L2 earth leakage current. The value must be less
than 1,000 microcamps amps.
1015010
BiPAP Vision Service Manual