Step 6
Again under Settings, open Communications then set the following:
Baud Rate; 19,200
Data Bits; 8
Stop Bits; 1
Parity; None
Flow Control; Xon / Xoff
Connector; COM 1 or COM 2 (which ever is being used).
Step 7
Click OK
Step 8
Turn on the Caps Lock.
Step 9
Enter the appropriate command for the test being performed. See the list of commands
on the next page.
Commands Options:
SJO
Transfer of total operating hours (MC subsystem connection).
SJB
Blower/valve calibration (PC subsystem connection).
SJL
System final test and limited PC function (PC subsystem connection).
SJP
PC error code info, including operating parameters (PC subsystem connection).*
SJM
MC error code information (MC subsystem connection).*
SJD
DC error code information (DC subsystem connection).*
SJE
Check Vent error code history (MC subsystem connection).
SJC
For determining type of breath triggering (MC subsystem connection).
* If the unit is in a Vent Inop condition, the current error will automatically appear.
BiPAP Vision Service Manual
Chapter 8: Testing and Calibration
8-39