5-14
Chapter 5: Routine Maintenance

5.9 Preventive Maintenance Schedule

The Maintenance Schedule lists the items that are recommended to be periodically inspected or tested. The
service interval may be decreased as internal protocol specifies. The user should be aware of any local or
national regulations that may deviate from the schedule as described below. Use the log to record the dates
the maintenance items are performed.
5.9.1 Vision Preventive Maintenance Schedule (
Model # ______________
Maintenance Item
Record hours of
operation
Replace inlet filter
Replace oxygen
regulator filter
Audible Alarm
Run Blower Valve
Calibration
Perform "System
Final Test"
Inspect Power
Cord
Cleaning
Tested by: ________________________________
1015010
Serial # ______________
Verification Reference Service Interval
Displayed on Options
Screen
Section 5.2
Section 5.4
Visual, verify by activating
Test Alarms.
Section 8.5
Section 8.8
Section 5.1
Section 5.7
BiPAP Vision Service Manual
Factory Recommended
1 Year
As required
As required
1 Year
1 Year
1 Year
As required
As required
Date: _______________
)