Chapter 8: Testing and Calibration

8.7 Run-In Cycle Procedure

Purpose
This procedure is to be used after servicing the BiPAP Vision as called out in the Recommended Testing after
Part(s) Replacement Chart in section 8.3.
parameters, to qualify the repair after a component has been replaced.
Equipment / Materials
• 0.25" test orifice (Respironics P/N 332353)
• pressure tubing
Procedure
Step 1
Connect the 0.25" test orifice (Respironics P/N 332353) to the outlet port, and the pressure
tubing from its port to the pressure input on the BiPAP Vision.
Connect the power, turn the unit On, and set up the following parameters in the S/T Mode.
Step 2
Step 3
Set the following alarm parameters:
If the unit has the optional Alarm Module, and/or optional Oxygen Module set the following
additional parameters:
1015010
IPAP = 40 cm H
O
2
EPAP = 4 cm H
O
2
Rate = 20 BPM
Timed Insp = 0.5 sec.
IPAP Rise Time = 0.1 sec.
Hi P = 50 cm H
O
2
Lo P = Disabled
Lo P Delay = 60 sec.
Apnea = Disabled
Lo MinVent = Disabled
Hi Rate = 50 BPM
Lo Rate = 4 BPM
%O
= 21%
2
BiPAP Vision Service Manual
The unit will cycle for one half hour, with specified operating