Respironics BiPAP Vision Service Manual page 205

Ventilatory support system
Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

A.2.4 PC/Laptop
Acceptable Options:
• Any commercially available PC or laptop with Windows operating system and hyperterminal.
A.2.5 Mechanical Lung (Adult)
Specifications:
Variable Compliance includes 0.05
Parabolic Resistor value RP-20
Lung capacity of at least 100L
Acceptable Options:
• Michigan Instruments, Inc. Model 1601 Test Lung
• Any commercially available test lung that meets the above specifications
A.2.6 Electrical Safety Analyzer
Specifications:
Earth Resistance
Range – 0 to 19.99 Ohms
Resolution – 0.01 Ohms
Accuracy – ±5%, ±1 digit
Leakage Current
Range – 0 to 19.99 µA
Resolution – 1µA
Accuracy – 2.5 to 1k H
1k H
100 k H
Acceptable Options:
•Dale model 544L
• Any commercially available safety analyzer that meets the above specifications.
: ±1%, ± 1 digit
Z
to 100 k H
: ± 1 digit
Z
Z
to 1M H
: ± 1 digit
Z
Z
BiPAP Vision Service Manual
Appendix A: Tools and Equipment
A-4