6.3.3.11 Modification of Parameter Values during the Centrifugation Run
(see 6.3.3.4 "Selection, Display, and Modification of Data"):
6.3.4
Program Mode
A program includes all of the data that are required for a
centrifugation run.
With a program, certain sedimentation results can be repeated under
identical conditions.
Programs can be loaded, executed, modified, and deleted at any
time.
A maximum of 50 programs can be stored under the numbers 1 - 50.
The rapid cooling program "RAPID_TEMP" does not occupy any
storage location and cannot be deleted. It is used to cool the
centrifuge without vessels.
"--" means that the values that are currently set are not a stored
program.
The programs can be protected against unauthorised use,
modification, or deletion with the aid of a code (see 6.3.3.7
"Configuration Menu" ).
Version 03/2010, Rev. 1.7 of 27/01/2012
Translation of the original operating manual
Modification possible
Speed
RCF
Runtime
Temperature
Acceleration curve
Deceleration curve
Attention: Temperature profiles cannot be created!
Modification
(De-)Activation
impossible
possible
Rotor
Continuous run
Program
Standstill cooling
Radius
Start delay
Density
Automatic lid opening
function
2-16KC
Using the Centrifuge
39