Error Codes - Sigma 2-16KC Operating Manual

Refrigerated centrifuge
Hide thumbs Also See for 2-16KC:
Table of Contents

Advertisement

2-16KC
Malfunctions ans Error Correction
7.2

Error Codes

Error no. Kind of error
1-9
System error
10-19
Speedometer
error
20-29
Motor error
30-39
EEPROM error
40-45
Temperature error
(only for refrigerated
centrifuges)
46-49
Imbalance error
50-59
Lid error
60-69
Process error
70-79
Communication
error
80-89
Parameter error
90-99
Other errors
Fig. 7.3: Error codes
If it is impossible to eliminate the errors, contact the service (see 0)!
44
Measure
allow to slow down
power off/on
allow to slow down
power off/on
power off/on
ensure ventilation
allow to slow down
power off/on
allow to slow down
power off/on
allow to cool down
provide better ventilation
allow to slow down
power off
eliminate the imbalance
press lid key
close lid
remove foreign matter from the
opening of the lid lock device
allow to slow down
power off/on
allow to slow down
power off/on
allow to slow down
power off/on
provide for better ventilation
check connections
Note
All these errors stop the centrifuge
or cause it to decelerate brakeless
With error 34, 35, 36, the
centrifuge will stop;
with error 37, 38 error message
only
With error 50 and 51, the
centrifuge will stop
With error 60 message "power
failure during run", with error 61,
message "stop after power on"
With error 83, error message only
Version 03/2010, Rev. 1.7 of 27/01/2012
Translation of the original operating manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

134226