25. WARNING AND ERROR MESSAGES

The following is a list of the error messages displayed by the SET and the
meaning of each message. If the same error message is repeated or if any
message not shown below appears, the instrument has malfunctioned. Contact
your Sokkia agent.
Bad condition
The air is shimmering a lot, etc., measuring conditions are poor.
The center of the target cannot be sighted.
Resight the target.
Unsuitable distance measurement conditions when reflectorless
measurement is set. When reflectorless measurement is set, distance cannot
be measured because the laser beam is striking at least two surfaces at the
same time.
Choose a single surface target for distance measurement.
Calculation error
Coordinates identical to the known point coordinates observed during
resection exist. Set another known point so that the known point coordinates
do not coincide.
Checksum error
A sending / repetition error has occured between the SET and external
equipment.
Send / receive the data again.
Communication error
A reception error has occurred in coordinate data from an external
instrument.
Check the settings of parameters concerning communication conditions.
"24.1 Changing Instrument Options
Flash write error!
Flash mount error!
It is impossible to read in data.
Contact your Sokkia agent.
Incorrect Password
Input password does not match set password. Input correct password.
Invalid baseline
During setting-out line measurement or point projection measurement,
baseline has not been defined correctly.
126