Memory is full
There is no more room to enter data.
Record the data again after deleting unnecessary data from the JOB or
coordinate data from the memory.
Need 1st obs
During missing line measurement, the observation of the starting position was
not completed normally.
Sight the starting position accurately and press [OBS] to perform the
measurement again.
Need 2nd obs
During missing line measurement, the observation of the target was not
completed normally.
Sight the target accurately and press [MLM] to perform the measurement
again.
Need offset pt.
Observation of the offset point during offset measurement was not completed
normally.
Sight the offset point accurately and press [OBS] to perform the
measurement again.
Need prism obs
During REM measurement, the observation of the target was not completed
normally.
Sight the target accurately and press [OBS] to perform measurement again.
New password Diff.
During new password setting, the passwords input twice are different. Input
new password twice correctly.
No data
When searching for or reading in coordinate data or searching for code data,
the search stopped either because the item in question does not exist or the
data volume is large.
No solution
The calculation of the instrument station coordinates during resection does
not converge.
Assess the results and if necessary, perform the observation again.