Part I: Connection and Filtration

1. Perform filling and venting as described in the previous chapter. If no pre-
use IT is needed the chapter "6 Wetting & Equilibration", page 25 can
be skipped.
2. Make sure the venting line valve (10) is closed.
3. Connect the process line to Inlet (3).
4. Connect the collecting line to Outlet (5). Make sure that the second out-
let option not connected to the collecting line is closed with a tube clamp.
5. Open inlet valve (8).
6. Open outlet valve (7).
7. Start the filtration process.
If required, please perform integrity testing of the device, following the same
instructions as for the pre-use integrity testing (see chapter "8.1 Integrity Test-
ing", page 33).
Instructions for Use Maxicaps
MR Virosart
HF, Max and Media
®
®
Filtration