Acknowledge Error Indication; Error - Cause - Remedy; Pop-Up Message - vacuubrand VACUU-SELECT Instructions For Use Manual

Hide thumbs Also See for VACUU-SELECT:
Table of Contents

Advertisement

Troubleshooting
View
message
Acknowledge
Tap/press
lightly
Error – Cause –
Remedy
80

8.1.2 Acknowledge error indication

8.1.2
Acknowledge error indication
Errors must be acknowledged after the fault has been remedied.
Error information and acknowledgement
1
Error message reset.
5
8.2 Error – Cause – Remedy
8.2
Error – Cause – Remedy
8.2.1 Pop‑up message
8.2.1
Pop‑up message
Error
Error
` Possible cause
Communication
One or more
`
error
VACUU·BUS compo‑
nents were removed.
Error at
Address incorrectly
`
frequency
configured.
converter (FC)
Temperature too high.
`
FC defective.
`
Error at control
Valve defective.
`
system
Error at pump
Check VMS‑B (switch‑
`
ing device).
Error at digital
No power supply at IN
`
I/O module
of I/O module.
Plug pulled out.
`
An error occurred in
`
the system and the I/O
module relayed it to the
controller.
2
Remedy
Deactivate relevant
VACUU·BUS compo‑
nents.
Perform component
detection.
Configure correct
address.
Replace defective
component.
Check address.
Replace defective
component.
Send in defective
device.
Connect power sup‑
ply.
Check plug‑in con‑
nection.
Remedy cause of
external error.
20901171_EN_VACUU·SELECT-Komp_V1.4_291021
Personnel
Specialist
Resp. spe-
cialist
Specialist
Resp.
specialist
Specialist,
resp.
specialist

Advertisement

Table of Contents
loading

Table of Contents