Troubleshooting
82
Error
Error
` Possible cause
No display
Power plug or plug‑in
`
power supply not cor‑
rectly plugged in or
pulled out.
Pumping unit switched
`
off.
VACUU·BUS plug‑in
`
connection or cables
defective or not con‑
nected.
Controller switched off
`
or defective.
Device fuse tripped.
`
Circuit board
Short circuit on the cir‑
`
fuse defective
cuit board.
Defective accessory
`
connected.
Power consumption too
`
high.
Transfer failed
No USB flash drive
`
connected.
Not enough storage
`
space on the USB flash
drive.
Venting valve
No voltage applied.
`
does not oper‑
VACUU·BUS plug‑in
`
ate
connection or cables
defective or not con‑
nected.
Venting valve dirty (pol‑
`
luted).
Venting valve in sensor
`
defective.
Venting valve deacti‑
`
vated.
No operation
Interface connected:
`
possible
Ethernet and/or
RS‑232.
Operation from external
`
terminal.
No license file
No USB stick inserted.
`
found
USB stick inserted wit‑
`
hout valid license.
Remedy
Check power connec‑
tion or plug‑in power
supply and cables.
Check VACUU·BUS
plug‑in connection
and cables to the
controller.
Replace defective
components.
Remedy cause of
the short circuit and
replace circuit board
fuse.
Send in.
Connect a USB flash
drive with sufficient
storage space.
Check VACUU·BUS
plug‑in connection
and cables to the
controller.
Clean venting valve.
If necessary, use
another external vent‑
ing valve.
Activate venting valve
in the controller.
Have operation
enabled from external
terminal.
Disconnect interface
connection.
Insert a USB stick
with a valid license.
20901171_EN_VACUU·SELECT-Komp_V1.4_291021