General Faults - vacuubrand VACUU-SELECT Instructions For Use Manual

Hide thumbs Also See for VACUU-SELECT:
Table of Contents

Advertisement

20901171_EN_VACUU·SELECT-Komp_V1.4_291021
Error
Error
` Possible cause
Error at analog
No power supply.
`
I/O module
Error at
Ambient temperature
`
Peltronic
too high, Peltronic over‑
heated.
Performance require‑
`
ments too high.
Peltronic defective.
`
Error at vacuum
Vacuum sensor defec‑
`
sensor
tive.
Overpressure
Pressure too high.
`
Measuring range
`
exceeded.
Underrange
Pressure below mea‑
`
suring range.
Vacuum sensor adjust‑
`
ment incorrect.
Maximum liquid
Full status indicator of a
`
level reached
level sensor.
Level sensor discon‑
`
nected.
Level sensor not adjus‑
`
ted correctly.
Component defective.
`

8.2.2 General faults

8.2.2
General faults
Error
Error
` Possible cause
Display frozen
Controller in undefined
`
state.
Controller has frozen.
`
Troubleshooting
Remedy
Connect power sup‑
ply.
Eliminate cause of
overheating of the
Peltronic.
Send in defective
component.
Replace defective
component.
Send in defective
component.
Acknowledge warning
indication.
Eliminate cause of
overpressure.
Calibrate vacuum
sensor correctly.
Empty the glass flask
or container in ques‑
tion.
Connect level sensor.
If permanently remo‑
ved, perform the
VACUU · BUS com‑
ponent detection.
Re‑adjust level sen‑
sor.
Exchange defective
component.
Remedy
Restart the controller.
Hold down ON/OFF
button for more than
10 seconds until
device reboots.
Personnel
Specialist
Specialist
Resp.
specialist
Operator,
specialist
Specialist
Operator
Personnel
Operator
81