vacuubrand CVC 3000 Instructions For Use Manual

Vacuum controller
Table of Contents

Advertisement

Technology for Vacuum Systems
Vacuum Controller
CVC 3000
Vacuum Controller
Instructions for use
Original instructions
N°: 999151_EN_PRINT

Advertisement

Table of Contents
loading

Summary of Contents for vacuubrand CVC 3000

  • Page 1 Technology for Vacuum Systems Vacuum Controller CVC 3000 Vacuum Controller Instructions for use Original instructions N°: 999151_EN_PRINT...
  • Page 2 +49 9342 808‑5550 ƒ Service +49 9342 808‑5660 Fax: +49 9342 808‑5555 E‑Mail: info@vacuubrand.com Web: www.vacuubrand.com Thank you for purchasing this product from VACUUBRAND GMBH + CO KG. You have chosen a modern and techni- cally high quality product. N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 3: Table Of Contents

    Product description 3.1 Vacuum controller CVC 3000 ..... . 20 3.2 Functionality ........22 3.3 Operation modes .
  • Page 4 5.2.4 Menu display in general ..... . 40 5.3 Handling CVC 3000 ......41 Operation 6.1 Switch‑on/‑off controller .
  • Page 5 10.1.1 Technical data......80 10.1.2 Product comparison CVC 3000 ....82 10.1.3 Rating plate.
  • Page 6 First steps First steps (delivery status) First steps on delivery status Select language and units N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 7: Introduction

    ƒ are intended to aid in your understanding of the proper use of the product. VACUUBRAND GMBH + CO KG reserves the right to modify ƒ or change the product design and/or technical specifications at any time without advanced notice.
  • Page 8: About This Document

    Please ask for replacement in case of an incomplete manual Contact us ƒ or download instructions for use on our web page: www.vacuubrand.com Contact us regarding any questions about this product, if you ƒ need further information, or to provide us with feedback.
  • Page 9: Symbols And Icons

    Introduction NOTICE Convention for additional notes Notice for a potentially harmful situation. Disregarding the notice could lead to material damage. Additional notes IMPORTANT! > Information or specific use recommendation, which must be observed. > Important information for the proper operation. Helpful tips and tricks >...
  • Page 10: Handling Instructions (Action Steps)

    Introduction Additional icons Positive example – Do! Negative example – References result – o. k. Do not! Refers to content of Refers to content of this manual. other documents. Handling or action Press selection knob or Turn selection knob key. Push and turn selection Push and hold key knob...
  • Page 11: Abbreviations

    Interface for an external peripheral device terms to connect with a VACUU·BUS capable ® vacuum gauge or ‑controller VACUU·BUS Bus system by VACUUBRAND. ® VACUU·CONTROL Web‑based application as remote control ® for vacuum controller and gauges made by VACUUBRAND. VACUU·LAN Local vacuum network for laboratories.
  • Page 12: Safety Instructions

    Use the product only when it is in proper working condition. 2.1.1 Intended use Intended use The Controller CVC 3000 is a laboratory instrument, used to measure and/or control vacuum in therefore intended plants. The controller may only be used in non‑explosive areas and indoors.
  • Page 13: Foreseeable Misuse

    To put the controller completely into vacuum. ■ To immerse the controller into liquid or to blast it with steam. ■ To use the remote control VACUU·CONTROL with ® CVC 3000 without knowledge of the connected vacuum sys‑ tem. N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 14: Target Groups

    Safety instructions 2.2 Target groups Ensure that the controller is only operated by authorized and IMPORTANT! skilled personnel. Users need to have the corresponding skills and qualifications for doing the job listed in the table User permissions. 2.2.1 User permissions This manual must be read, understood and complied with by the person performing one of the following tasks: Supervising...
  • Page 15: Safety Precautions

    Safety instructions 2.3 Safety precautions Products of VACUUBRAND GMBH + CO KG are subject to high Quality standards and safety quality tests with goals for safety and operation. Prior to delivery each product has been tested thoroughly. 2.3.1 Protective clothing...
  • Page 16 >Repairs may only be performed by qualified personnel, e. g., service technician. CVC 3000 in combination with VACUU·CONTROL ® Option VACUU·CONTROL is a remote control for the controller. It is ®...
  • Page 17 ATEX category and peripherals controller need to have the same or must have a higher ATEX approval. Without concordant categorization of peripherals, the specified category of the VACUUBRAND equipment loses its validity. Avoid ignition sources The use of gas ballast or the operation of venting valves is only...
  • Page 18: Safety And Service

    Safety instructions 2.4 Safety and service Safety regulations that apply to your work environment also apply Obligations for to persons who perform service works, especially in the handling service jobs of hazardous materials. 2.4.1 Meaning Health and Safety Clearance Products which are potentially hazardous may only be returned when all dangerous contaminations are removed.
  • Page 19: Environmental Protection

    ƒ media). Any more questions? We will help you: Phone: +49 9342 808‑5660 Fax: +49 9342 808‑5555 service@vacuubrand.com 2.5 Environmental protection NOTICE Risk of environmental damage due to incorrect disposal of the controller. Do not dispose your product in household waste! >...
  • Page 20: Product Description

    999151 Safety Information for Vacuum Equipment 999254 Origin packaging ‑‑‑‑‑‑ 3.1 Vacuum controller CVC 3000 The controller is designed for applications requiring controlled vacuum. The controller has a two‑point control mode to switch an in‑line isolation valve. The controller is freely programmable. Up to 10 programs can be stored in the controller memory.
  • Page 21 Product description Front side Front panel Meaning Chemically resistant plastic housing LC display Control panel and product name Selection knob Rear side Rear side Meaning Ports for VACUU·BUS components ® Mains connection ` power supply unit ` VARIO diaphragm pump or ®...
  • Page 22: Functionality

    Product description 3.2 Functionality The controller manages vacuum processes by controlling vacuum Functionality pumps, in‑line isolation‑ and/or air admittance valves. It controls process vacuum, cooling water and venting to demand. Valves and/or vacuum pumps are necessary to operate the controller. Without those components the controller can only be used as vacuum measurement device.
  • Page 23: Operation Modes

    Product description 3.3 Operation modes Up to 5 different operation modes are selectable at the controller. Specific modifications can be realized by indivudual mode me‑ nus. Selectable operation modes Pump down Standard ƒ Vac control ƒ Program ƒ Auto mode Optionally ƒ...
  • Page 24: Installation And Connection

    Installation and connection Installation and connection The controller is designed for installation directly at the workplace. > Observe all specifications for installation, connection and operation according to technical data,  see chapter 10.1.1 Technical data. > Also observe rating plate data. Installation conditions The controller has acclimatized.
  • Page 25: Built-In Version

    Installation and connection 4.1.2 Built‑in version The CVC 3000 can be used as built‑in version fixed with spring clips; e. g., as front controller of a VARIO pump, built into a cable ® duct cover or into the cut‑out of a switch cabinet.
  • Page 26: Examples Of Use

    Vacuum hose Rotary evaporator  Example CVC 3000 build‑up with cabinet dryer Wall power supply Controller CVC 3000, table top version Vacuum sensor (VSK) Cabinet dryer with control unit on top Vacuum valve Vacuum hose Diaphragm pump, vacuum pump N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 27 Installation and connection  Example CVC 3000 directly installed Wall power supply Controller CVC 3000, directly mounted VACUU·BUS cable ® Vacuum vessel, recipient Vacuum valve Vacuum hose Diaphragm pump, vacuum pump Install the controller as close as possible to the process in >...
  • Page 28: Connection

    Installation and connection 4.2 Connection 4.2.1 Electrical connection Wall power supply kit* Power supply kit for CVC 3000 * short-circuit-proof multi-voltage power supply with integrated overload protection and changeable mains plugs. Prepare wall power supply plug 1. Take the wall power supply kit out of the packaging.
  • Page 29 > Plug female connection of the power supply cable into mains connection of the controller. Mains connection on the rear side Consider new connection design: For easy connection, the CVC 3000 Ports with guide of the newest series have a guide groove groove on the rear side for each port.
  • Page 30: Vacuum Connection

    Installation and connection 4.2.2 Vacuum connection NOTICE Flexible vacuum hoses can contract because of evacuation. Fix vacuum hose at the connections. > > Fix connected components. > > Measure and trim the vacuum hose to a length that cares for >...
  • Page 31 Direct installation Flange connection (4) at vacuum chamber with sealing ring (5) ƒ ƒ and small flange (6) screwed to CVC 3000 fixed with clamping ring (7). Built‑in version (front mounting) Vacuum hose made of PTFE (9) – plugged on hose nipple, ƒ...
  • Page 32: Vacuu·bus

    Y adapters make it possible to connect up to 32 accessory components. VACUU·BUS principle ®  Example Principal sketch 1 CVC 3000 4 Vacuum sensor VSK VACUU·BUS ® components 2 Y adapter 5 Vacuum valve (In‑line valve) 3 Extension cable 6 Level sensor When connecting multiple, identical VACUU·BUS...
  • Page 33 Installation and connection VACUU·BUS components ® VACUU·BUS ® Vacuum sensor VSK 3000 636657 accessories VSP 3000 636163 (Option) Vacuum gauge VACUU·VIEW 683220 VACUU·VIEW extended 683210 Vacuum valve VV‑B 6 674290 (In-line valve) VV‑B 6C 674291 VV‑B 15C, KF 16 674210 VV‑B 15C, KF 25 674215 Coolant valve...
  • Page 34: Operating And Display Elements

    5.1 Operating elements The operating elements are located on the controller front. controller figure  see chapter 3.1 Vacuum controller CVC 3000 on page 20 5.1.1 Selection knob The selection knob of the controller is a combination of rotary knob and push‑button.
  • Page 35: Key Combinations

    Operating and display elements – Select operation mode Mode Control panel keys ` With stopped operation: Mode menu for selecting the operation mode. – Change function/mode Mode ` During running operation: To switch from Pump down and further to Vac control Auto mode.
  • Page 36: Display And User Interface

    After booting the pressure display appears, including Bar graphic and preset operation mode. 5.2.1 Pressure display  Example Display after switching on CVC 3000 Title bar (or status bar) Meaning Operation mode ‑ ` Pump down Mode ` Vac control...
  • Page 37: Display Icons

    Operating and display elements 5.2.2 Display icons When vacuum control has started additional icons appear on the display. When starting operation Icon Meaning Display symbols during operation Vacuum control is running (animation) Process time; runtime vacuum control (hh:mm:ss) Active component Icon Meaning Icons for active...
  • Page 38 Operating and display elements Status display while operation is running Icon Meaning Icons for control Pump down – continous pumping status Pump down: lower pressure limit reached VACUU·LAN: pump down to set pressure Vac control: for 2‑point control – pump down to set pressure VACUU·LAN: pressure increase to switch on pressure Vac control: preset maximum exceeded VARIO control: pump down to set point.
  • Page 39: Signal Sounds (Warning Beep)

    Operating and display elements Fault indication (warning symbol) Icon Meaning Error display Flashing: warning! When device connected to SUB‑D (Option) Icon Meaning Active connection to Controller in remote mode; controller only controllable RS232 port via connected PC or Notebook, local operation disabled. VACUU·CONTROL ‑adapter connected;...
  • Page 40: Menu Display In General

    Operating and display elements 5.2.4 Menu display in general The controller includes several menus and submenus, e. g., Configuration, Function, Display..Submenu  Example Submenu Configuration Meaning Titel bar ` Menu name Menu interface ` Selectable functions or submenus (on the right side of the display) ` Adaptable value or ` selectable content Back (return)
  • Page 41: Handling Cvc 3000

    Operating and display elements 5.3 Handling CVC 3000 The handling of the controller is menu‑driven. Menus are ac‑ Handling and cessed via push buttons on the control panel or via key combina‑ Operation tions. Use the selection knob to select function or menu.
  • Page 42 Operating and display elements Input (data entry) Changeable values are positioned on the right side in the display. Text on the ride side accords to content selection like in a drop down list. Exception: menu Program, in this menu data and value are editable. Example: enable entry and edit 1.
  • Page 43: Operation

    Operation Operation 6.1 Switch‑on/‑off controller Switch‑on Switch on CVC 3000 CVC 3000 V2.xx 5 Initial screen: company logo and firmware version, for approximately 2 seconds. 5 Pressure graphic is displayed. Switch‑off > Press key Switch‑off CVC 3000 On/Off 5 Controller switched off (display off).
  • Page 44: Select Language And Pressure Unit

    Operation 6.2 Select language and pressure unit Select language and pressure unit 櫓匡 櫓匡 CVC 3000 V2.xx 5 Initial screen. N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 45: Mode - Operation Mode

    Operation 6.3 Mode – Operation mode The controller is supplied with several operating modes. Only when the controlling process is stopped it is possible to select an operation mode. 6.3.1 Select operation mode  Example Call‑up Mode menu 5 Title bar shows the selected operation mode (Mode). Select any other operation mode in the same way as described above for control.
  • Page 46: Mode Menu

    Operation 6.3.2 Mode menu Mode menu display  Example Menu description Title bar menu name Meaning and operation modes Selectable operation modes Pump down ` Continous pump down or ` Pump down with pressure and time presetting. ` VARIO : pump down with adjustable motor speed (pumping ®...
  • Page 47: Start Controlling

    Start vacuum controlling after selecting the required operation mode. The controller works in delivery status with the default set‑ tings of the factory setting. Start controlling  Example Start CVC 3000 5 Controller starts. 5 Icons are displayed. 6.5 Control during operation 6.5.1 Venting (VENT)
  • Page 48 Operation Venting VENT button is used to vent the system. A short click on this button will momentarily vent the system as the process contin‑ ues. Holding the VENT key for longer than 2 seconds will cause the system to be vented to atmospheric pressure and the pump will stop running;...
  • Page 49: Change Operation Mode

    Operation 6.5.2 Change operation mode During running operation the operation mode can be switchted between by pressing Pump down, Vac control Auto mode Mode key. Switch mode during running operation  Examples Switch mode Pump down ↓ Vac control 5 Operation mode switched to control.
  • Page 50 Operation Switch back from Vac control Pump down. Switch back to primary mode 5 Title bar displays Pump down. Typical applications Pump down control:  Semiautomatic distillation. Recommended for applications for which the process vacuum is still to be determined. Firstly the vacuum pump is pumping down rapidly in mode Pump down.
  • Page 51: Display Graphic (Curve)

    Operation 6.5.3 Display graphic (curve) In addition to the bar graphic of the pressure display the display Pressure history Graphic can be switched to a diagram named which shows a pressure vs. time curve. That Graphic curve will only be displayed while operation is run‑ ning.
  • Page 52: Quit Display Graphic

    Operation Graphic menu Description pressure history display Header Meaning ` Active operation mode. ` Vacuum set point; for Vac control Auto mode. ` Actual vacuum value (actual pressure). ` Eelapsed process time. Axis – pressure ` Unit according to pre‑settings (mbar, Torr, hPa) Pressure graph ` Pressure/time progress Axis –...
  • Page 53: Quick Adaption During Operation

    Operation 6.6 Quick adaption during operation 6.6.1 Set vacuum Vac control In mode the set vacuum can be adapted directly du‑ ring running operation. Adapt set vacuum  fine tuning 1 detent = 1 pressure value (mbar, Torr, hPa)  Example Adapt set vacuum fine tuning 5 Controller controls to new set vacuum.
  • Page 54: Motor Speed (Only Vario ® )

    5 Pump icon with percentage value. 5 VARIO pump runs with adjusted motor speed. ® 6.7 Stop control Stop control  Example Stop CVC 3000 5 Controller and vacuum control stops. 5 Display icons switched off. N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 55: Advanced Menus And Operation

    Advanced menus and operation Advanced menus and operation 7.1 Operation menus A selected operation mode can be adapted and optimized for the Optimizing operation process through the corresponding operation menu. The settings mode in an operation menu include mainly: motor speed, set vacuum or time presettings.
  • Page 56: Pump Down

    Advanced menus and operation 7.1.1 Pump down Continous pump down with pressure and time presettings. Meaning Menu – Pump down  Example Operation menu Pump down Parameter Meaning Parameter Pump down Speed (%) VARIO : Speed settings for pump down. ®...
  • Page 57: Vac Control

    Advanced menus and operation 7.1.2 Vac control Control to a set vacuum value. Meaning Menu – Vac control  Example Operation menu Vac control Parameter Meaning Parameter Vac control Set vacuum Setting for lower vacuum level for 2‑point control (mbar) or precisely for VARIO pump.
  • Page 58: Auto Mode

    Advanced menus and operation Hysteresis values Auto Factory settings Set vacuum 900 1000 (mbar) Hysteresis Hysteresis (mbar) 7.1.3 Auto mode (only valid for VARIO pump) ® Automatic detection and tracking of boiling point, unaffected by Meaning varying process conditions. Menu – Auto mode ...
  • Page 59: Program

    Advanced menus and operation Application example – rotary evaporator Minimum to prevent rotary evaporator from receiving re‑eva‑ Application example poration from the flask. Set Minimum value to the vapor pressure for Auto mode of the solvent at ambient temperature. 7.1.4 Program Up to 10 individual programs including vacuum and time preset‑...
  • Page 60: Vacuulan

    Advanced menus and operation 7.1.5 VACUULAN Vacuum control, optimized for VACUU·LAN vacuum networks ® Meaning by VACUUBRAND. Menu VACUULAN  Example Display VACUULAN Parameter Meaning Parameter VACUULAN Set vacuum Setting for lower vacuum level, which shall be reached easily when the vacuum network is not (mbar) used.
  • Page 61: Configuration Menu

    Advanced menus and operation 7.3 Configuration menu In menu Configuration the controller parameters are set. This Meaning menu is also for adjusting the vacuum sensor and for loading Defaults settings. Call up Configuration menu  Example Call up Configuration menu Configuration menu displayed.
  • Page 62: Content Selection

    Advanced menus and operation 7.3.1 Content selection Configuration The following menu items of can be selected, ac‑ Specified content tivated and used. Defaults Selection Meaning Defaults (standard) Cancel Leave menu item without default setting. Load Load default settings. Defaults Load is activated all controller parameter will be IMPORTANT! reset to delivery status.
  • Page 63: Submenus

    Advanced menus and operation Adjustment Selection Meaning Sensor adjustment function 1060–700 Adjustment range of a vacuum sensor, internal or ex‑ 20–0 ternal at atmospheric pressure (1060–700) or under vacuum (20– ~0). For further descriptions about sensor adjustment  see chapter 9.2 Sensor readjustment on page 77 7.3.2 Submenus Submenu –...
  • Page 64 Advanced menus and operation Submenu – Sensors In submenu Sensors all connected sensors are listed. The inter‑ nal sensor is generally displayed as Sensor. External sensors are listed with sensor type name and address. Submenu Sensors Display Meaning Sensor selection Inverse Sensor = currently selected for pressure display.
  • Page 65 Advanced menus and operation Submenu – RS‑232 Submenu RS-232 is applied for interface configuration, parame‑ ter adjustments and commands.  see also Online Instructions for use. Submenu RS232 Parameter Selection Meaning Adjustable RS232 parameter Baud 19200 Default setting for transmission speed. The baud rate of data transfer of trans‑...
  • Page 66: Function Menu (See Online Version Of Cvc Manual)

    Advanced menus and operation Function menu (see Online version of CVC manual) Program functions (see Online version of CVC manual) Detailed descriptions about function menu with address assign‑ ment or about differential pressure measurment  see Online Instructions for use 999151.
  • Page 67: Resolving Problems

    Resolving problems Resolving problems Technical support > To identify errors and potential remedies, please refer to the Technical support troubleshooting table: Fault – Cause – Remedy on page 69 In case you need additional assistance, please contact our Service department. 8.1 Error display The major symbol for fault indication is the warning triangle.
  • Page 68 Resolving problems Combinations of flashing display icons Icon flash rate Fault and Meaning beep when Sound ` Limit pressure reached ` Overpressure ` Process time elapsed ` Venting valve ` In‑line suction valve ` Coolant valve ` External sensor removed ` or defective ` internal sensor defective ` Vario pump...
  • Page 69: Fault - Cause - Remedy

    Resolving problems 8.2 Fault – Cause – Remedy Fault ` Possible cause Personnel 3Remedy User, Sensitive process not Motor speed too high. Reduce motor speed. ƒ ƒ Specialist controllable Pumping speed too high. ƒ respon. Frequent error messa‑ Several controllers are Use only one controller for ƒ...
  • Page 70 Resolving problems Fault ` Possible cause Personnel 3Remedy User, In‑line suction valve icon In‑line suction valve Check the connection. ƒ ƒ Specialist flashes removed. Check plug connection. ƒ Plug disconnected. ƒ Replace defective parts. ƒ In‑line suction valve ƒ Reconfiguration without In‑ ƒ...
  • Page 71 Resolving problems Fault ` Possible cause Personnel 3Remedy Specialist No display Controller switched off. Switch on the controller; ƒ ƒ On/Off key. Power supply disconnec‑ ƒ ted. Check plug connection ƒ and wall power supply for Power supply not cor‑ ƒ...
  • Page 72 Resolving problems Fault ` Possible cause Personnel 3Remedy User, Sensors submenu is per‑ Submenu Sensors... Select the required sensor ƒ ƒ Specialist manently displayed does not automatically by turning and pressing switch back to previous selection knob. display. respon. After loading defaults Special factory settings Set language and pressure ƒ...
  • Page 73: Controller Reset

    Resolving problems 8.3 Controller Reset Auto reset The following error indications will be reset automatically with Automatic reset remedy: Overpressure ƒ Process time elapsed ƒ Limit pressure reached ƒ Error air admittance valve ƒ Error Peltronic ƒ Active reset Several error indications need to be reset manually. Depending Reset after action on the fault severity different actions are required.
  • Page 74 Resolving problems IMPORTANT! Note the settings of stored programs, before loading Default. > > Load Default settings Load factory settings N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 75: Error Of External Components

    Resolving problems Load factory settings 8.4 Error of external components Error messages for defective external components such as In‑ line suction valve, vacuum sensor, etc. cannot be reset. > Replace defective accessories or > send defective accessories for repair to your local supplier or to our Service.
  • Page 76: Cleaning And Maintenance

    Cleaning and maintenance Cleaning and maintenance 9.1 Cleaning This chapter does not contain descriptions for the deconta‑ IMPORTANT! mination of the controller. This chapter describes only simple cleaning and care measures. 9.1.1 Controller Clean surface > Clean soiled surface with a clean, slightly wetted cloth. To moisten the cloth we recommend water or mild soap.
  • Page 77: Sensor Readjustment

    Cleaning and maintenance 9.2 Sensor readjustment NOTICE For readjustment the reference pressures need to be known with certainty In the pressure range 20 – 700 mbar (15 – 525 Torr) no adjustment is possible. Check the accuracy of the pressure sensor in case of irre‑ >...
  • Page 78 Cleaning and maintenance Adjustment under vacuum An adjustment under vacuum is only possible if the pressure is Adjustment under vacuum lower than < 20 mbar (< 15 Torr) absolute. 1. Evacuate the measurement connection of the controller or in case the connected external vacuum sensor VSK 3000 to a pressure <...
  • Page 79 Cleaning and maintenance Adjustment at a reference pressure Instead of adjustment under vacuum to a pressure < 0,1 mbar Adjustment at (< 0.1 Torr), adjustment to a precisely known reference pressure reference pressure within the range of 0 – 20 mbar (0 – 15 Torr) is possible. 1.
  • Page 80: Appendix

    108 mm x 58 mm x 34 mm 4.3 in. x 2.3 in. x 1.4 in. Weight 300 g 0.66 lb Mains plug AC, changeable: EU/UK/US/AUS Electrical data – CVC 3000 (US) Supply voltage, max. 24 VDC (±10 %) 24 VDC (±10 %) Power, max. 3,4 W 3.4 W...
  • Page 81 DN 6/10 Venting Hose nozzle for flexible tube = 4–5 mm Vacuum data (only differing data) Technical data package for fine CVC 3000 with external VSP 3000 (US) vacuum control Measuring range, absolute 1000–1x 10 mbar 750–1x 10 Torr ‑3...
  • Page 82: Product Comparison Cvc 3000

    5 in. x 5 in. x 3.5 in. Dimensions with foot 144 mm x 124 mm x 115 mm 6 in. x 5 in. x 4.5 in. * without wall power supply 10.1.2 Product comparison CVC 3000 Product comparison CVC 3000 Product name CVC 3000 ‑...
  • Page 83: Rating Plate

    Service department, name us product > > type and serial number. With this information we can offer selective support and advice for your product. Rating plate CVC 3000 Rating plate Manufacturer Type/Date of construction/Month Serial number Power supply Electrical power Compatible to VACUU‑BUS...
  • Page 84: Interface Commands (See Online Version Of Cvc Manual)

    Detailed descriptions about the complete interface commands  see Online Instructions for use 999151. 10.2.1 Pin assignment (RS232) Sub‑D panel connector SUB‑D 9 poles Sub‑D 9 poles (rear side of CVC 3000) PIN Name Operation PIN Name Operation ‑‑‑ ‑‑‑ Received data...
  • Page 85: Ordering Information

    Vacuum controller Order‑N° Ordering information CVC 3000 table top version 683160 CVC 3000 built‑in version 636595 CVC 3000 + VSP 3000 (package for fine vacuum control) 635983 Spare parts Order‑N° Ordering information spare parts Vacuum sensor (vacuum gauge head) VSK 3000, 1080‑0,1 mbar...
  • Page 86 Purchase original accessories and spare parts from your spe‑ International sales offices and cialized distributor or through international sales offices of specialized trade VACUUBRAND GMBH + CO KG. Information about the complete product range are available > > in the current product catalog.
  • Page 87: Service

    Appendix 10.4 Service Take advantage of the comprehensive service range of Service range VACUUBRAND GMBH + CO KG. Service in detail Product guidance and practical solutions, ƒ fast delivery of spare parts and accessories, ƒ professional maintenance, ƒ immediate repairs processing, ƒ...
  • Page 88: Glossary

    It is operated by one or two vacuum controllers CVC 3000. If two CVC 3000 are connected to the VMS‑B it switches off the pump only if both applications do not need a vacuum supply any‑...
  • Page 89: Index

    ....35 Menu VACUULAN ..60 CVC 3000 ....Specification and features . . . 22 Mode.
  • Page 90 VACUULAN ... . 46,60 Vacuum connection ..30 Vacuum controller CVC 3000 . 20 vacuum port ....30 VACUU·VIEW.
  • Page 91: Certifications

    Bevollmächtigter für die Zusammenstellung der technischen Unterlagen / Person authorised to compile the technical file / Personne autorisée à constituer le dossier technique: Dr. J. Dirscherl · VACUUBRAND GMBH + CO KG · Alfred‑Zippe‑Str. 4 · 97877 Wertheim · Germany Ort, Datum / place, date / lieu, date: Wertheim, 20.07.2017 ppa.
  • Page 92: Us/Can Certificate

    Appendix 10.7.2 US/CAN Certificate N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 93: Declaration Of Conformity - China Rohs 2

    10.7.3 Declaration of Conformity – China RoHS 2 DECLARATION OF CONFORMITY – China RoHS 2 VACUUBRAND GMBH + CO KG has made reasonable efforts to ensure that hazardous mate‑ rials and substances may not be used in its products. In order to determine the concentration of hazardous substances in all homogeneous materials of the subassemblies, a “Product Conformity Assessment”...
  • Page 94 (Pb), mercury (Hg), cadmium (Cd), hexavalent chromium (Cr+VI), polybrominated biphenyls (PBB), and polybrominated diphenyl ethers (PBDE). Products manufactured by VACUUBRAND may enter into further devices (e.g., rotary evaporator) or can be used together with other appliances (e.g., usage as booster pumps).
  • Page 95 N°: 999151_EN_CVC3000_PRINT/20.07.2017...
  • Page 96 Technology for Vacuum Systems Manufacturer: VACUUBRAND GMBH + CO KG Alfred‑Zippe‑Str. 4 97877 Wertheim GERMANY Phone: ƒ Head office +49 9342 808‑0 ƒ Sales +49 9342 808‑5550 ƒ Service +49 9342 808‑5660 Fax: +49 9342 808‑5555 E‑Mail: info@vacuubrand.com Web: www.vacuubrand.com...

Table of Contents