Installation and connection

Check incoming
goods
Check installation
conditions
IMPORTANT!
30
4
4
Installation and connection
Installation and connection

4.1 Transport

4.1
Transport
Products from VACUUBRAND are packed in sturdy, recyclable
packaging.
The original packaging is accurately matched to your
product for safe transport.
If possible, please keep the original packaging, e.g., for
>
returning the product for repair.
Goods receipt
Check the shipment for transport damage and completeness.
> Immediately report any transport damage in writing to the
supplier.
> Compare the scope of delivery with the delivery note.

4.2 Installation

4.2
Installation
Check installation conditions
The device is acclimatized.
ƒ
Ambient conditions have been observed and are within the lim‑
ƒ
itation of use.
Limitation of use
Ambient temperature
Max. altitude
Relative humidity
Protection class (front)
Prevent condensation or contamination from dust, liquids, or corrosive
gases.
> Note the IP protection class of the controller.
> IP protection is only guaranteed if the controller is
appropriately mounted or installed.
10‑40 °C
2000 m
above sea level
30–80 %, non‑condensing
IP 42/IK 06