NOTE
NOTE
Condensate can damage the electronics.
A large temperature difference between the storage location
and the installation location can cause condensation.
After goods receipt or storage, allow your vacuum device to
>
acclimatize for at least 3‑4 hours before initial use.

4.2.1 Benchtop version

4.2.1
Benchtop version
As a benchtop version, the controller can be set up directly on the
work surface and connected, for example on the lab bench.
4.2.2

4.2.2 Lab scaffold version

Lab scaffold version
With the lab scaffold version, a stand holder is fitted to the back
of the controller. The stand holder enables the controller to be at‑
tached directly to a scaffold bench in the lab or mounted on the
wall using the wall bracket.
Installation and connection
31