Connect Unit To Electrical Power; Attach Power Cord - Thermo Scientific Revco ULT1790-10-A Operating Manual

Cxf -40c and -86c chest freezers
Table of Contents

Advertisement

Section 1
Installation and Start-up
Remote Alarm Contacts
and Analog Output (cont.)

Attach Power Cord

Connect Unit to Electrical
1-10
Revco CxF
The analog output function allows the
freezer to output signals representing
the temperature of the freezer cabinet.
The factory default setting is 4-20 mA.
Refer to Table 1-2 for output
specifications.
Table 1-2. Analog Output Specifications
Temperature
Insert the power cord into the power inlet module. Place the retaining
bracket (P/N 195763) over the connector. Tighten retaining screws to
secure.
Figure 1-12. Power Cord Attachment
Note See the serial tag on the side of the unit for electrical specifications or
refer to the electrical schematics in this manual.
Power
The freezer should be operated on a dedicated, grounded service. Check
the voltage rating on the serial tag of the unit and compare it with the
outlet voltage. Then, with the power switch turned off, plug the line cord
into the wall outlet.
First, turn on the freezer power switch. Then open the lower front door by
grasping the bottom left corner. Locate the battery switch (Figure 1-4) and
turn it to Standby mode ( ). During initial freezer start-up, the system
battery may require charging and the Low Battery message may appear in
the message center.
Note Ensure the battery switch is turned to Standby mode ( ). The
rechargeable batteries require 36 hours to charge at initial start-up. A "Low
Battery" alarm may occur until the batteries are fully charged. Should a
power failure occur during the initial start-up period, the electronics will
have limited operation.
4-20 mA
0-1V
-100 to +50°C
-100 to +50°C
s
I I M M P P O O R R T T A A N N T T U U S S E E R R I I N N F F O O R R M M A A T T I I O O N N
CAUTION! Stored product should be protected
by a redundant 24 hour/day monitoring system
with alarm capability. An interconnect jack and
thermocouple are installed for centralized
monitoring, should on-board system fail.
0-5V
-100 to +50°C
s