4.2

Master/Slave board specifications

Master/Slave board
Outputs
Inputs
Master/Slave configuration Star: one master mainframe can drive up to eight slave main-
Cabling
Connectors
Synchronization
Accuracy
Distance
Indicators
36
8, one master mainframe can drive up to eight slave main-
frames
1, combined with master output
frames in a star configuration. No daisy-chaining
Fiber-optic
Fiber-optic LC-type connectors
Clock (timebase), trigger, qualifier, acquisition
Built-in delay measurements confine accuracy within ± 50 ns
Up to 300 m between master and slave
2 LED indicators per channel:
one for connection status and one for data
I2710-1.0 en