Performance Verification; Dc Accuracy Test - Teledyne T3HVD Manual

Voltage probe, high voltage differential probes
Table of Contents

Advertisement

Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Storage Altitude
Pollution Degree

Performance Verification

The table below lists the test instruments required for the performance verification process.
Oscilloscope
Signal Generator /
Calibrator
Digital Multimeter
Pincer Clips
BNC Adapters
Connect power to the probe and allow 30 minutes warm up time.

DC Accuracy Test

Connect the T3HVD probe BNC output to the Digital Multimeter input using the Male BNC to Binding
Posts adapter. Connect the male BNC connector to the T3HVD female BNC output. Connect the
Multimeter test leads to the binding posts.
Connect the T3HVD input leads to the calibrator output directly or using the pincer clips.
Set the probe sensitivity to the lower voltage range, apply the DC voltage from the calibrator according to
the table below, and record the output voltage on the DMM. Repeat for the higher voltage
T3HVD1500-70,
T3HVD1500-200
T3HVD7000-100
Page 11
Accuracy < 1.5%, e.g. Teledyne LeCroy WaveSurfer
4000HD or HDO4000A Series
Amplitude Accuracy < 0.75%, Rise Time < 3ns, e.g.
Fluke / Wavetek 9100
6.5 digits minimum, e.g. Teledyne LeCroy
T3DMM6-5, Keithley 2000
Supplied in probe accessories
BNC Male to BNC Female T Piece, BNC Male to
Dual Binding Posts, BNC Female to Dual Binding
Posts, BNC Male to
50 x
5 V
500 x
50 V
100 x
10 V
1000 x
100 V
o
o
0
C - 50
C
o
o
-30
C - 70
C
< 85% Relative Humidity
< 90% Relative Humidity
3000 m
12000 m
2, Indoor Use Only
50Ω Termination
100 mV + 2 mV
100 mV + 2 mV
100 mV + 2 mV
100 mV + 2 mV
Displays probe output
Bandwidth, AC
Accuracy, CMRR
DC Accuracy
General Testing
General Testing
range.