Checking The X-Ray Tube Assembly; Checking The Tube Current - Dentsply Sirona Heliodent Plus Installation Instructions Manual

Hide thumbs Also See for Heliodent Plus:
Table of Contents

Advertisement

8 Starting up
8.1 Functional test
70
8.1.4

Checking the X-ray tube assembly

1. Switch on the building installation or the emergency circuit to which
the Heliodent Plus was connected.
2. Switch the unit on.
3. Monitor the self-test.
4. Select the shortest radiation time (0.01 - 0.05) and 60kV.
WARNING
X-ray radiation!
5. Release radiation 3 times.
8.1.5

Checking the tube current

Auxiliary devices required
● Ammeter
CAUTION
Only use battery-powered measuring devices.
Preparation
1. Turn the unit off.
2. Set the ammeter to the "10 mA DC" measuring range.
3. Connect the ammeter to the two outer X600 connectors on board
DX1.
Dentsply Sirona
Installation Instructions Heliodent Plus
62 14 972 D 3507
D 3507.031.01.15.02 03.2018