Final Work - Dentsply Sirona Heliodent Plus Installation Instructions Manual

Hide thumbs Also See for Heliodent Plus:
Table of Contents

Advertisement

9 Final work
80
9

Final work

1. Switch the unit on.
2. Finally, check the functioning of the unit by releasing radiation
repeatedly with different settings (time, kV, tooth symbol, patient
symbol) according to the operating instructions.
The unit must function flawlessly.
3. Instruct the customer on how to operate the unit.
The unit is now ready for operation.
➢ For the USA/Canada only: Attach the warning label and the DHHS
label.
Dentsply Sirona
Installation Instructions Heliodent Plus
62 14 972 D 3507
D 3507.031.01.15.02 03.2018