Maintenance Instructions

Tube Current Verification

Turn unit OFF and remove housing (see Service Manual!).
Connect multimeter to X600, range 10mADC.
Turn unit ON. (WARNING: Electrical shock hazard!)
Set the exposure time on the control panel to 3.2s, using the + or - button
(conventional radiography technique).
CAUTION RADIATION: Make an exposure.
The reading should be 7mADC ±1.4 mA. Record reading.
If specified value is obtained turn unit OFF.
Remove meter leads!
– If specified value cannot be obtained, see Service Manual, chapter "Tube Current Verification".
6
Sirona Dental Systems GmbH
7 mA ±1.4 mA
62 15 144 D 3507.101.01.03.02