Checking the X-ray tube assembly joint

CAUTION
Nature and source of the danger
Switch OFF the unit before connecting a measuring instrument or
replacing parts!
Required tools
Torx screwdriver (size 10, 15 and 20)
Check
● Does the cone remain set in every position?
● Are the connecting cables intact?
– See section on "Checking the connection cables".
If the X-ray tube assembly fails any of these check points, it must be
replaced (see Section "Replacing X-ray tube assembly H1 [ → 81]").
Checking the connection cables
1. Switch the X-ray unit off.
2. Unscrew and remove the old cover (A).
3. Remove the arm cover (L).
L
4 Maintenance
4.8 Checking the X-ray tube assembly joint
55