ASK Rate

Generate ASK Modulation Using SCPI Commands

U2761A User's Guide
The ASK rate is the rate at which the output amplitude
"shifts" between two preset amplitudes. The range of the
ASK rate is 2 mHz to 100 kHz. The default ASK rate value is
10 Hz.
Soft Front Panel Operation
On the ASK Rate panel as shown in
desired ASK rate value and select the unit from the drop
down list.
Remote Interface Operation
ASKey:INTernal:RATE <rate in Hz>
Example 1
-> *CLS; *RST
-> ASK:STAT ON
-> FUNC SIN
-> FREQ 1000
-> VOLT 5 VPP
-> VOLT:OFFS 0
-> ASK:INT:RATE 10
-> OUTP ON
Features and Functions
Figure 2-
26, input the
//To reset the U2761A to
default power-on state, this
command can be ignored if
this operation is not
required.
//Enables ASK modulation.
//Sets the carrier waveform to
Sine wave.
//Sets the carrier frequency to
1 kHz.
//Sets the output amplitude to
5 Vpp.
//Sets the output offset to 0.
//Sets the ASK rate to
10 Hz.
//Turns on output.
2