2
The duty cycle setting does not apply to a Square waveform
used as the modulating waveform for AM, FM, or PM. A 50%
duty cycle is always used for a modulating Square waveform.
The duty cycle setting applies only to a Square waveform
carrier.
Soft Front Panel Operation
After selecting Square
at the top panel as shown in
Figure 2-
2, on the Duty Cycle panel below, input the desired
duty cycle value.

Figure 2-8 Panel view of the duty cycle section

Remote Interface Operation
FUNCtion:SQUare:DCYCle <percent>
The APPLy command automatically sets the duty cycle to
50%.