Carrier Frequency; Table 2-4 Carrier Frequency For Fm - Agilent Technologies AGILENT U2761A User Manual

Usb modular function/arbitrary waveform generator
Table of Contents

Advertisement

Carrier Frequency

U2761A User's Guide
Remote Interface Operation
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset.
The maximum carrier frequency depends on the selected
function as shown below. The default is 1 kHz for all
functions.

Table 2-4 Carrier frequency for FM

Function
Sine
Square
Ramp, Triangle
Arbitrary
The carrier frequency must always be greater than or equal
to the frequency deviation plus 1 μHz. If you attempt to set
the deviation to a value plus 1 μHz greater than the carrier
frequency (with FM enabled), the U2761A will automatically
adjusts the deviation to the maximum value allowed with the
present carrier frequency.
The sum of the carrier frequency and deviation must be less
than or equal to the maximum frequency for the selected
function (20 MHz for Sine and Square, 200 kHz for Ramp
and Arbitrary waveforms). If you attempt to set the
deviation to a value that is not valid, the U2761A will
automatically adjusts it to the maximum value allowed with
the present carrier frequency.
Features and Functions
Minimum frequency
1 μHz
1 μHz
1 μHz
1 μHz
2
Maximum frequency
20 MHz
20 MHz
200 kHz
200 kHz