Example 1
-> *CLS; *RST
-> FUNC SIN
-> VOLT 5 VPP
-> VOLT:OFFS 0
-> SWE:STAT ON
-> FREQ:STAR 500
-> FREQ:STOP 10000
-> SWE:SPAC LIN
-> SWE:TIME 2
-> TRIG:SOUR IMM
-> OUTP:TRIG ON
-> OUTP:TRIG:SLOP POS
-> OUTP ON
//To reset the U2761A to
default power-on state,
this command can be
ignored if this operation
is not required.
//Sets the carrier waveform
to Sine wave.
//Sets the output amplitude
to 5 Vpp.
//Sets the output offset to
0.
//Enables frequency sweep.
//Changes the start
frequency to 500 Hz.
//Changes the stop
frequency to 10 kHz.
//Changes the sweep mode
to Linear.
//Changes the sweep time
to 2 s.
//Changes the sweep trigger
source to Internal.
//Turns on the trigger
output.
//Outputs the trigger on a
rising edge.
//Turns on output.