6. TROUBLESHOOTING

PROBLEM
GREEN LIGHT ON, THE
POWER SWITCH OFF
ANOMALOUS LEVER
MOVEMENTS (STEPS
MOVEMENTS)
IT'S NO POSSIBLE TO
EXECUTE A PM OR
EXCENTRIC EXERCICE, THE
DEVICE DON'T START
THERE IS AN OFFSET ON
COUPLE VALUE
NO CONNECTION
BETWEEN DEVICE AND PC
Easytech s.r.l. – Genu Iso/Plus User's Manual
CHECK TO BE PERFORMED
Is there electric power on the
plug where the device is
installed?
Is the power cable correctly
connected on the device? Is it in
good condition?
Is the main fuse good?
Low battery.
Exhausted Batteries.
Air may be present in the
hydraulic circuit
Low battery
Exhausted Batteries.
It's necessary to reset the
torque offset.
Control the emergency stop
button, it must be lifted
Control the USB cable, it must
be well inserted in the plug
Maybe it's necessary restart the
device and the PC
rev. C
SOLUTION
- Contact our electrician.
- Connect with care the power cord
on the three-pole plug on the device.
Control the power cable and
eventually change it.
- Checks the main fuse and replace if
necessary, as described in cap. 4
- Leave the device connected to grid,
with the main switch in ON position,
some hours without use it.
- Change the batteries.
- Follow the air elimination described
in the cap. 4.4.2
- Leave the device connected to grid,
with the main switch in ON position,
some hours without use it.
- Change the batteries.
Follow the procedure described in the
cap 4.4.3
Put the button in lifted position, as
indicated by the arrows
Connect with care the USB cable
Switch off the device and the PC.
Execute this operation following the
order: control and eventually
reconnect the USB cable – switch on
the PC (leave the operating system
start) – switch on the device – start
the program
Pag. 80 di 97