T CaRe Compact
Mod. 80000650
User's Manual
EASYTECH s.r.l.
Via della Fangosa, 32 – 50032 Borgo San Lorenzo (FI) – Italy
tel. 055 8455216 – fax 055 8454349
e-mail:
info@easytechitalia.com
Easytech technical support
Direct line of
:
+39 348 2316323
1936
Easytech s.r.l. –T CaRe Compact User's Manual
rev. D
Pag. 1 di 36

Advertisement

Table of Contents
loading

Summary of Contents for easytech T CaRe Compact

  • Page 1 Via della Fangosa, 32 – 50032 Borgo San Lorenzo (FI) – Italy tel. 055 8455216 – fax 055 8454349 e-mail: info@easytechitalia.com Easytech technical support Direct line of +39 348 2316323 1936 Easytech s.r.l. –T CaRe Compact User’s Manual rev. D Pag. 1 di 36...
  • Page 2 Version 02/03/2020 Title: T CaRe COMPACT User’s Manual Internal Revision MusTCareCOMPACT _GB_BN_ D (Based on MusTCareCOMPACT_IT_BN_C) Easytech s.r.l. –T CaRe Compact User’s Manual rev. D Pag. 2 di 36...
  • Page 3: Table Of Contents

    MALFUNCTIONS ........................29 MAINTENANCE ........................30 Cleaning the exterior ..................... 30 Periodic inspection and life cycle of the unit ............30 Calibration and settings (by the Easytech Technical Service) ......31 Fuses replacement ......................31 TRANSPORT ..........................32 8.1. Environmental and condition for transportation and storage ......32 8.2.
  • Page 4: Introduction

    1.2 Applied standards The device T CaRe Compact is made in accordance with Directive 93/42/EEC and subsequent amendments, applying the current version of the standards listed below: ...
  • Page 5 – Part 1: General Requirements  EN 1041 Information supplied by the manufacturer of medical devices  ISO 13485 Medical devices – Quality management systems – Requirements for regulatory purposes Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 5 di 36...
  • Page 6: Prescriptions For Safety

     For information concerning electromagnetic interferences, refer to paragraph 2.5. Avoid in particular installation of the device close to potential sources of electromagnetic Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 6 di 36...
  • Page 7: Precautions For Use

     If the device has fallen to the ground or the external casing is damaged or you reasonably suspect that the performance of the device has degraded, disconnect it from the power Easytech supply and contact the technical support of Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 7 di 36...
  • Page 8: Warnings Regarding Electromagnetic Interference

    Therefore, there are no essential "therapeutic" performances to be maintained in critical situations. WARNING T Care Compact is a Class A product. In a residential environment this device may cause radio interference in which case the user may be required to take adequate countermeasures.
  • Page 9 Guidance and manufacturer's declaration – Electromagnetic Emissions [tab.1/60601-1-2] T CaRe Compact is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Emission test...
  • Page 10 3 Veff/m2 from 80MHz to 80MHz to 2700MHz 2700MHz 27 Veff/m2 from 27 Veff/m2 from 380MHz to 380MHz to 390MHz 390MHz 28 Veff/m2 from 28 Veff/m2 from Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 10 di 36...
  • Page 11 Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 Veff/m Tabella 2-3 – EMC, Immunity Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 11 di 36...
  • Page 12 Recommended separation distances between portable and mobile RF communications equipment [tab.6/60601-1-2] T CaRe Compact is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of T CaRe Compact can help prevent electromagnetic...
  • Page 13: Special Prescriptions For The Operator

     patients with acute skin diseases;  diabetic patients with vascular problems. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 13 di 36...
  • Page 14: Particular Caution

    The device is also equipped with safety devices for automatic overcurrent protection (fuses), which act independently of the control software and are able to suspend the operation of the entire apparatus. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 14 di 36...
  • Page 15: Description Of The Device

    The unit is equipped with 2 electrode holders and one standard reference electrode support placed on the bottom of the device. The power switch is located on the side panel; on the same side there are also two fuses. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 15 di 36...
  • Page 16: Electrodes

    The three standard electrodes provided with all machines are shown in figures 3.2, 3.3, 3.4.The figure 3.5 show a couple of elbow adapters (optional) useful for permit a different positioning of the electrode (cap and res). Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 16 di 36...
  • Page 17 Figure 3-5 Elbow pipes - The optional electrodes are shown in the figures 3-6, 3-7, 3-8; Figures 3-6 – Functional Resistive Electrode Figures 3-7 – Return Plates Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 17 di 36...
  • Page 18 Easytech invites to use the type of cream that comes with the device, tested and EC marked. The cream is made with standard ingredients tested in applications identical to the present.
  • Page 19: Sales Configuration

    Capacitive Electrodes (complete with cover) D. 75 – D 55 – D 35 n°1 Return Plate n°1 Conductive Cream 100 ml n°2 Adhesive Reference plate 136 cm n°1 Plastic Case with foam padding n°1 Backpack Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 19 di 36...
  • Page 20: T Care Compact Technical Characteristics

    Dimensions WxDxH 225x300x185 mm Weight 4,4 Kg Environment temperature + 10 ÷ + 40°C Relative humidity (non-condensing) 0 ÷ 75% Atmospheric Pressure 700 ÷ 1060 mbar Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 20 di 36...
  • Page 21: Using T Care Compact

    On the knob there is also a key for command. 4.2 Operations with user interface Easytech When the device is switched-on, for about 2 seconds the logo and indication CaRe Compact are displayed. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 21 di 36...
  • Page 22: Treatment

    Figure 4-1- Manual setting of operating parameters – Time mode Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 22 di 36...
  • Page 23 In reality the value of the voltage supplied by the device is programmed; for simplicity, it is called power. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 23 di 36...
  • Page 24 Aesthetic), to the resistive electrode and the electrode Face (only Physio Aesthetic) the LED colour is orange-red, for those capacitive and Body (Physio Aesthetic) you have a blue light. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 24 di 36...
  • Page 25: Set-Up Commands

    The vertical arrows are used for selecting a row, the horizontal ones to change a parameter (press to confirm the change, to exit). The modified parameters are stored in non- volatile memory, so they are valid for subsequent use. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 25 di 36...
  • Page 26: Guidelines For Treatment With T Care Plus

    Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 26 di 36...
  • Page 27: Operating In Resistive Mode

    2) Scimitar: For transverse and detachment mobilization maneuvers 3) Dragon: For muscle-joint treatment 4) Dragon II: For the treatment of large area, including the spine 5) Dolphin: For myofascial treatments Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 27 di 36...
  • Page 28: Operating In Capacitive Mode

    If this condition occurs, the treatment should be discontinued and the electrode should be smeared with conductive cream and possibly repositioned. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 28 di 36...
  • Page 29: Malfunctions

    Easytech For any real or suspected malfunction, contact Technical Support (Easytech Contact Details are shown on the cover page of this User Manual). Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 29 di 36...
  • Page 30: Maintenance

    Easytech ensures the maintenance of the unit up to 5 years after its exit from the market (according to existing law). Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 30 di 36...
  • Page 31: Calibration And Settings (By The Easytech Technical Service)

    7.3 Calibration and settings (by the Easytech Technical Service) Calibration consists of a procedure that allows to evaluate the error of the machine on the measure of a variable whose value is significant for the purposes of treatment, in order to determine whether to accept it (as negligible) or to correct it through calibration or to repair if it is a failure.
  • Page 32: Transport

    9. UNIT DISPOSAL AT THE END OF ITS LIFECYCLE The device no longer usable must be disposed in accordance with current law provisions concerning electric and electronic equipment, WEEE Directive 2012/19/UE. Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 32 di 36...
  • Page 33: Labels

    Production year The device that is no longer usable must be disposed of in accordance with current regulation for electrical and electronic equipment (WEEE Directive 2012/19/UE) Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 33 di 36...
  • Page 34 Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 34 di 36...
  • Page 35 Easytech s.r.l. – User Manual T CaRe Compact rev. D Pag. 35 di 36...
  • Page 36 EASYTECH s.r.l. Via della Fangosa, 32 – 50032 Borgo S. Lorenzo Firenze – Italy Tel.: +39 055 8455216 – Fax: +39 055 8454349 E-Mail: info@easytechitalia.com www.easytechitalia.com Easytech technical support Direct line of +39 348 2316323...