Power frequency
magnetic field (50/60
Hz)
IEC 61000-4-8
Guidance and manufacturer's declaration – Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment.
Immunity Test
RF Conducted
from 150 KHz to
IEC 61000-4-6
80 MHz
6,765-6,795 MHz
13,553-13,567 MHz
26,957-27,283 MHz
40,66-40,70 MHz
RF radiated
80MHz ÷ 800 MHz
IEC 61000-4-3
(See note)
800MHz ÷ 2.7GHz
In the equation, "d" is the recommended separation distance (in meters) and "P" is the maximum rated output
power of the transmitter (in Watt) according to the manufacturer; "E" is the level of immunity; "*" is the
multiplication symbol, "
Interference may occur near devices marked with the following symbol:
Easytech s.r.l. – Genu Iso/Plus User's Manual
30 A/m
Compliant
Conformity
Test Level
3 Veff
(Standard A)
6 Veff
3 V/m
(Standard A)
3 V/m
(Standard A)
" means square root.
Distanza dalla sorgente di disturbo (m) raccomandata
Portable and mobile RF communication
equipment should be used no closed to any
part of the device, including cables, than the
racommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
d = (6/E) * P
d = (6/E)* P
d = (6/E) * P
d = (6/E) * P
rev. C
Pag. 11 di 97