2.5 SPECIAL PRESCRIPTIONS FOR OPERATORS

There are no specific prescriptions for operators. If an operator wishes to test the
operation of the device on his own body, he must simply follow the prescriptions for
patients.
During the different phases of the exercises, the operator must:
Supervise the exercise in order to be able to intervene in case of need.
Use the device in accordance with the instructions contained in this manual and
avoid performing any operation that could endanger himself and/or the patient.
Verify that all safety distances are maintained during the active phase of the
treatment in order to prevent potential collisions between the patient and other people
or objects.
Help elderly patients and patients that have difficulties in moving or that are
unstable.
Disconnect the device from the power supply and contact the technical support of
Easytech if the connection cable is damaged or if liquid has penetrated in the device.

2.6 SPECIAL PRESCRIPTIONS FOR THE PATIENT

2.6.1 Contraindications

At present, Genu/Prima can be used by anyone provided that the physician/therapist
defines the type, intensity and limits of the exercises in function of the ability and
conditions of each patient.
The apparatus of the Isokinetic Line must not be used on pediatric patients.

2.6.2 Special cautions

The are no particular precautions to be taken in the use of the device, unless it
verifies the correct positioning of the patient.

2.7 SAFETY DEVICES

If a power failure occurs on the 220V mains, when power resumes the device sets itself
in standby mode and waits for the operator to select a function.
The device is also fitted with automatic safety devices that protect it from overcurrrents
(fuses) and that are able to stop the device independently from the software.
It is provided an additional thermal control that stops the engine when the
internal temperature goes too high.
Easytech s.r.l. – Genu Iso/Plus User's Manual
The patient has available an emergency button that instantly cut the power to
the electrical power of the appliance, making it inert. If the emergency button
is pressed a message will appear on the screen of the PC (no for concentric
mode)
rev. C