4. CONNECTION TO GRID, FIRST TURN ON AND START
PROGRAM
4.1 CONNECTION TO GRID
Insert the power cord to device and to grid (100-240 V 50/60 Hz)
Make sure the Emergency Stop Button is in a raised position, if not, put it in its fully
raised position by turning it as shown by the arrows.
Turn on the device by using the green switch, it will light up.
CHECKS TO BE PERFORMED IF THE DEVICE WILL NOT POWER ON
Is the grid power present in the device's
installation site?
Is the power cord correctly connected to
device? Is it intact?
Is the main fuse good?
Easytech s.r.l. – Genu Iso/Plus User's Manual
CHECK
SOLUTION
- Contact your own electrician
- Connect the power cord whit care to
device. Check the power cord and
eventually replace it.
- Check the main fuse and replace if
necessary.
rev. C
Pag. 18 di 97