4.4.3

Procedure for torque offset reset

In case an offset is noticed on torque indication, it is possible to reset the torque value.
Enter the setup following the steps indicated in the previous paragraph and click on
"0Nm" button and following the instructions on the screen.

4.4.4 Notice recovery of configuration files

There is a possibility that a configuration file error might occur when the program
starts up. An automatic file retrieval mechanism has been implemented, the program
retrieves a copy of the saved configuration file at the end of the previous working
session and restores it successfully.
Through the following message the program notifies the operator of the error
resolution.
Click the "OK" button and proceed with a normal working session.
If the error persists, contact our service center.