In the following paragraph, every time the expression generic movement is used
(with no reference to a specific limb), it refers to the muscles and movements of
the knee, that is to a flexion/extension movement.
Selection of the LIMB - Move the mouse over the figure of the human body and
click the desired limb.
Setting the ANGLE – This phase is used to calibrate the encoder. Position the
patient correctly (as instructions in section 4.3.1), attaching the lever to the ankle
using the system provided. Fully extend the leg and press the 0° button to save the
position. Eventually it is possible to do the same thing for 90° (bent leg). It is
advisable to set the parameters when the leg is extended (0°), unless the therapist
judges that it is impossible.
Number of repetitions – Move the cursor until the desired number displays and
move the mouse to the right or left in increments of units for a fine adjustment.
Extension/flexion speed - Set the desired value by moving the cursor. Move the
cursor to the right or left in increments of units for a fine adjustment.
SYNCRO – The extension/flexion values will be the same.
.
To move to the next step, select LIMITS SETTING or COMPENSATION
5.3.1.2 Setting the limits
1
2
3
4
Easytech s.r.l. – Genu Iso/Plus User's Manual
rev. C
Pag. 33 di 97