Use
20 mm
20 mm
20 mm
1.0 Nm
50.0 %
54
Live view
-20 mm
20 mm
20 mm
1
2
The illustration shows how the alignment recommenda
tion for the frontal view from the sample instructions for
use for the prosthetic knee joint can be readjusted and
checked with several auxiliary lines in the live view.
1.0 Nm
50.0 %
: "Done" – 1x tapping closes the pop-up for the auxili
ary line
: "0 mm" – The list with the reference numbers of
prosthetic knee joints for which the value for the align
ment reference point in the sagittal view is stored is
available here.
: "±" – 1x tapping positions the auxiliary line on the
opposite side of the load line – this function is important
when the list with reference numbers of prosthetic knee
joints is used, since the alignment in the live view is
opposite depending on the amputation side, and has to
be checked and corrected as needed
: "- 0 mm +" – Change the value by 1x tapping of +
and - or by entering the value
Drawing auxiliary lines
2x tapping on the load line opens the popup for the
auxiliary line – a maximum of three auxiliary lines can be
positioned on the load line. In the sagittal view, this is
only possible on the load line facing the camera.
Tapping and dragging can be used to position the
auxiliary line.
743L500 3D L.A.S.A.R. Posture