Missing Ground Contact; Calibration Failed - Otto Bock 743L500 3D L.A.S.A.R. Posture Instructions For Use Manual

Hide thumbs Also See for 743L500 3D L.A.S.A.R. Posture:
Table of Contents

Advertisement

Use

5.5.3 Missing ground contact

5.5.4 Calibration failed

40
Error message text:
: "Missing ground contact"
: "A force sensor is not placed on the floor."
: "Please check the surface below the measurement
plate." – Apply the load to the force measurement plate
as evenly and consistently as possible. Avoid a point
load near the edge of the plate since this can also
cause this error message.
: "OK"
Possible causes:
Uneven floor
Defective sensor in the force measurement plate
Defective circuit board in the force measurement
plate
Measures:
► Carry out the appropriate measures depending on
the cause (for example level the force measurement
plate on an even floor; have defective components
repaired by Ottobock Service).
► 1x tap on "OK".
Calibration in this section deals with the cameras' abil­
ity to detect the position of the force measurement
plate.
Error message text:
: "Calibration failed"
: "The calibration failed"
: "Cancel"
: "Recalibrate"
Possible causes:
One or more LEDs are not recognised, are covered
or are outside the outer blue frame
The camera stand or force measurement plate was
moved during the calibration
Overexposed image (white)
Underexposed image (black)
Backlighting
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog­
nised within the outer blue frame; restart the mini-
computer in case of incorrect exposure).
Take corresponding steps to prevent backlighting
► 1x tap on "Recalibrate".