User Maintenance
Assembling Check

Short Plate Check

Measuring the fixture's impedance with the short plate in place should check
the condition of the short plate.
"Short Plate Check"
been completed.
Required Tools
Table 4-14
Setting of Measuring Instrument
Measurement Condition
Electrical Length
Measurement Parameter
OSC Level
Point Averaging
To make the same settings as given in the above table, refer to the
Operation Manual for the measuring instrument.
Procedure
Step 1. Clean the short plate as described in
Step 2. Remove the cap and place the short plate with the protruding surface down
on the insulator assembly.
Step 3. Place the cap on the fixture body and fasten it.
Step 4. Measure Rs and Ls at 100MHz and 1GHz in this condition.
Keysight 16196A/B/C/D Parallel Electrode SMD Test Fixture
should be performed after the
1.5-mm hex wrench (provided accessory)
Impedance measuring instrument (with 7-mm connector and calibrated)
"Electrode Check"
Set Value
16196A: 26.2mm
16196B: 26.9mm
16196C: 27.1mm
16196D: 27.3mm
Ls, Rs
500mV
32
"Cleaning" on page
34.
has
49