Service
Replacement Procedure
Upper Electrode
1. Prepare the replacement electrode.
2. Remove the cap from the fixture.
3. Take out the screw from the top of the cap and remove the electrode from
In case of the 16196D, take out the screw from the top of the cap while
holding the electrode.
4. Take the electrode out of the spring and insert the replacement electrode.
5. Push the electrode in from the bottom of the cap so that the top of the
Figure 6-7
Electrode Replacement 1
72
the cap.
electrode protrudes out of the top of the cap.
Keysight 16196A/B/C/D Parallel Electrode SMD Test Fixture