Cleaning Methods - Keysight 16196A Operation And Service Manual

Table of Contents

Advertisement

User Maintenance
Cleaning

Cleaning Methods

Upper Electrode (
Open Plate, Short Plate
Use the Cleaning Rod (Keysight parts number 5182-7586) for the cleaning.Use
the white rubber part of the cleaning rod to remove dirt from all contacting
surfaces of the above-mentioned parts. Be careful not to scratch or damage
the parts when removing the dirt.
Figure 4-3
Cleaning Rod
Dirt tends to adhere to the upper electrode and short plate parts in particular.
Meticulous cleaning of these parts is recommended.
The front of the short plate has a sharp edge, so take adequate precautions
when cleaning it.
Do not use a file or similar object to remove dirt, as this will affect
measurement accuracy and repeatability.
If the dirt cannot be removed, replace the part. For replacement method,
see the sections
Lower Electrode (
Use the Cleaning Rod (Keysight parts number 5182-7586) for the
cleaning.First use the push ring to press down the insulator assembly. While
maintaining this condition, use the white rubber part of the cleaning rod to
remove dirt from the contacting parts of the lower electrode. Be careful not to
scratch or damage the parts when removing the dirt.
Dirt tends to adhere to the lower electrode parts in particular. Meticulous
cleaning of these parts is recommended.
Do not use a file or similar object to remove dirt, as this will affect
measurement accuracy and repeatability.
Keysight 16196A/B/C/D Parallel Electrode SMD Test Fixture
(1)), Insulator Assembly (
Figure 4-2
"Parts Replacement"
and
(3))
Figure 4-2
Figure 4-2
"Procedure for
Replacement".
(4)),
35