Operation

Selecting and Changing the Insulator Assembly

Selecting and Changing the Insulator Assembly
Select an insulator assembly that corresponds to the shape of the DUT being
measured and replace the insulator assembly in the fixture.
An exclusive type of insulator assembly is supplied with each model. Do
not use an insulator assembly from a different model.
Step 1. Select an insulator assembly that is appropriate for the shape of the DUT to
be measured.
To take accurate and repeatable measurements, it is necessary for the DUT to
be placed in the DUT insertion hole and be stable. For that reason, the
16196A and 16196B each are provided with 3 types of insulator assembly
which have DUT insertion holes with different diameters (the 16196C has
only one type of insulator assembly, and the 16196D has two types of
insulator assembly). Select an insulator assembly that will create the
narrowest gap between the DUT and the DUT insertion hole.
Figure 3-1
DUT and DUT Insertion Hole Diameter
Table 3-1
Insulator Assembly Specifications
16196A
20
Insulator
Example of Corresponding Chip (mm)
Assembly
Length (L)
1.34
1.6
1.14
1.6
1.08
1.6
Keysight 16196A/B/C/D Parallel Electrode SMD Test Fixture
Width (W)
Height (H)
0.8
0.8
0.8
0.6
0.8
0.5