Rs-485-Related Setting Values; I/F Setting; Id/ Slave Address For Modbus-Rtu; Communication Type - Unipulse F701-P Operation Manual

Table of Contents

Advertisement

9 RS-485 INTERFACE
9-3.

RS-485-related setting values

■RS-485 I/F setting
■RS-485 ID/ slave address for MODBUS-RTU
Input the ID setting (setting mode 5-3)

■Communication type

Set the RS-485 operation. (setting mode 5-2)
Chapter
9
56
56
1.
Set the RS-485 port of this device. (Setting mode 5-1)
* If the communication type is Modbus-RTU, set as length of character: 8 bits and stop bit:
1 bit (stop bit: 2 bits if the parity bit is none).
RS-485 I/F setting
Baud rate
0: 1200bps
1: 2400bps
2: 4800bps
3: 9600bps
4: 19200bps
5: 38400bps
The length of character
0: 7bit
1: 8bit
2.
Make initial settings of the RS-485 port of the personal computer, PLC, etc.,
connected according to the settings of this device.
RS-485 ID (slave address)
Communication type
Mode
0: Command
1: Auto
2: Continuous
3: Modbus-RTU
Terminator
(effective at only UNI-Format)
0: CR
1: CR+LF
Stop bit
0: 1bit
1: 2bit
Parity bit
0: None
1: Odd
2: Even
ID setting (0 - 31)
Format
(effective at "auto" or "continuous")
0: Gross weight
1: Net weight
2: Gross weight + Net weight