FUNCTION SETTINGS

5-1.

Display update rate (setting mode 4-5)

Set the number of updates of the indicated value per second. What is selected here is only the display
update rate. The internal A/D conversion rate and CPU processing speed stay unchanged.
The display update rate can be selected from 1 time/sec, 2 times/sec, 5 times/sec, 10 times/sec, and 20
times/sec. Normally, use 20 times/sec.
If the indicated value flickers too much to read at 20 times/sec, decrease the display update rate.
* When the temperature is low, may automatically adjust the display update rate.
5-2.

Subdisplay selection (setting mode 4-5)

Set the data displayed in the subdisplay area.
The data set here are displayed in the subdisplay area.
The contents of display are any of the following: Accumulation Count/Accumulation Value,
Accumulation Value, Accumulation Count, Final, and none.
(For the contents of display in detail, see ② Subdisplay under "■Front panel" on page 2.)
5-3.

Digital low pass filter (setting mode 2-1)

This low pass filter cancels undesired noise components by filtering the A/D-converted data.
Set the cutoff frequency like a low pass filter in an analog circuit.
The cutoff frequency can be selected from 1.5, 2, 2.5, 3, 4, and 5Hz. Select an optimum value
according to the type of weighing and setting environment.
31
Display selection
Display selection
5 FUNCTION SETTINGS
Display update rate
0: 1 times/sec.
1: 2 times/sec.
2: 5 times/sec.
3: 10 times/sec.
4: 20 times/sec.
Subdisplay selection
0: Accumulation Count/Accumulation Value
1: Accumulation Value
2: Accumulation Count
3: Final
4: None
Chapter
5
31