Span Calibration (Setting Mode 9-1) - Unipulse F701-P Operation Manual

Table of Contents

Advertisement

4 CALIBRATION
4-5.

Span calibration (setting mode 9-1)

Put a balance weight on the load cell (scale), and register the span (gain).
- Put a balance weight of the weight set by balance weight value on the load cell (scale).
- Check for unwanted loads as in the case of zero calibration.
- Check that "
Chapter
4
Setting mode No.
* If an error message is displayed, see "11-2.Over scale/error display" on page 100.
30
30
(Calibration with a balance weight of 50% or more of the capacity is advantageous in terms of
linearity, etc.)
" is on.
(When the indicated value is not stable, accurate calibration cannot be performed.)
1.
Select setting mode 9-1 (span calibration).
2.
Perform span calibration.
Press the
key and start the registration of balance weight value.
Upon completion of input, fix it with the
3.
Span calibration ends with the weight value display equal to the balance weight
value.
Weight value
key.
Display of span calibration in progress.
kg
kg

Advertisement