Calibration Error Display - Unipulse F701-P Operation Manual

Table of Contents

Advertisement

11 SUPPLEMENTS

■Calibration error display

Chapter
11
The initial tare exceeds the zero calibration range of the F701-P. Check the load cell for
unnecessary load.
If "
the +EXC and -SIG terminals of the load cell and redo zero calibration after shifting the zero point.
The relationships between connected resistance values and input signals are shown below.
- The numerical values in this table assume that one 350Ω load cell is used.
- The temperature coefficient of the resistance connected here directly affects the accuracy of the
102
102
(Calibration error)
" is displayed under a normal load, there is a need to connect resistance between
+EXC
+SIG
-EXC
-SIG
Resistance value
Approximate value
Calculated value
875  kΩ
437  kΩ
291  kΩ
219  kΩ
175  kΩ
146  kΩ
125  kΩ
109  kΩ
97  kΩ
87.3 kΩ
79.4 kΩ
72.7 kΩ
67.1 kΩ
62.3 kΩ
58.2 kΩ
54.5 kΩ
51.3 kΩ
48.4 kΩ
45.9 kΩ
43.6 kΩ
41.5 kΩ
39.6 kΩ
37.9 kΩ
36.3 kΩ
34.8 kΩ
indicator.
Use resistance of 50ppm/ C or higher (approx. 5ppm/ C is recommended).
Connect resistance
between +EXC and -SIG.
μ-STRAIN
866  kΩ
442  kΩ
294  kΩ
221  kΩ
174  kΩ
147  kΩ
124  kΩ
110  kΩ
97.6 kΩ
86.6 kΩ
78.7 kΩ
73.2 kΩ
66.5 kΩ
61.9 kΩ
57.6 kΩ
54.9 kΩ
51.1 kΩ
48.7 kΩ
46.4 kΩ
43.2 kΩ
41.2 kΩ
39.2 kΩ
38.3 kΩ
36.5 kΩ
34.8 kΩ
Input conversion strain
mV/V
200
0.1
400
0.2
600
0.3
800
0.4
1000
0.5
1200
0.6
1400
0.7
1600
0.8
1800
0.9
2000
1.0
2200
1.1
2400
1.2
2600
1.3
2800
1.4
3000
1.5
3200
1.6
3400
1.7
3600
1.8
3800
1.9
4000
2.0
4200
2.1
4400
2.2
4600
2.3
4800
2.4
5000
2.5