11 SUPPLEMENTS
11-2.
Over scale/error display
Chapter
11
■Over scale display
The weight error output turns ON.
A/D converter input over
Net weight > Net over setting value
Gross weight > Capacity + 9 scale divisions
Gross weight > Gross over setting value
* Net weight = Gross weight - Tare
■Calibration error display
The initial tare exceeds the zero adjustment range.
The initial tare is minus (negative).
The span setting value is larger than the capacity.
The span setting value is "00000".
The load cell (scale) output has not reached the span adjustment range.
The load cell (scale) output is minus (negative).
The load cell (scale) output exceeds the span adjustment range.
The weight value is not stable and calibration is stopped.
100
100
Error description